4 Comments

  1. James on October 4, 2018 at 3:28 pm

    This is a great idea. So useful. Please keep the posts coming.

  2. Matamis Dimitrios on October 8, 2018 at 3:12 am

    Reverse triggering or excessive inspiratory effort beyond the mechanical inspiratory time

    • Thomas Piraino on October 10, 2018 at 12:15 pm

      Your second comment was correct. The reason why it is not reverse triggering is that the breath is patient triggered as demonstrated by the pressure deflection prior to breath delivery. This same looking waveform can occur with reverse triggering, but there would be no indication of patient trigger.

  3. Juan Martín Núñez silveira on October 14, 2018 at 11:20 pm

    It s a premature cycling by increased drive